Development of an automatized system for a gamma scanning technique

Main Article Content

Francisco Rodríguez-Méndez
Marcela Meneses-Guzmán
Bruno Chinè-Polito
Roberto Pereira-Arroyo

Abstract

Non-destructive testing techniques have proven to be a powerful tool in the study and characterization of materials. This work presents the development of an automated measurement system based on the attenuation of gamma radiation for the study of materials. First, we described the design process of the electromechanical elements and the automatic control system, followed by its implementation. The results show the capacity and mechanical accuracy of the system. The resolution of the motor and the displacement capacity of the robot are 8 μm for the horizontal movement with maximum carriage travel of bi-directional 54 cm and 4 μm for vertical movement, with maximum carriage travel of bi-directional 57 cm. The system allows for a sample rotation, being the angular trajectory of 360 ° with unidirectional travel. The sensitivity of the system for detecting specific changes in the thickness of the material is verified statistically, also evaluating the sampling periods to obtain a measurement with precision. The relative error of the measurements is determined and its causes and solutions are discussed.

Article Details

How to Cite
Rodríguez-Méndez, F., Meneses-Guzmán, M., Chinè-Polito, B., & Pereira-Arroyo, R. (2019). Development of an automatized system for a gamma scanning technique. Tecnología En Marcha Journal, 32(3), Pág. 87–98. https://doi.org/10.18845/tm.v32i3.4482
Section
Artículo científico

References

[1] M. I. Haraguchi, H. Y. Kim, F. E. Sprenger y W. A. P. Calvo, «Industrial Equipment Troubleshooting with Imaging Technique Improved Gamma-Ray Absorption Scans,» Journal of Physical Science and Application, vol. 9, nº 2, pp. 359-371 , 2012.
[2] «ASTM E1316-17 Standard Terminology for Nondestructive Examinations,» ASTM International, West Conshohocken, PA, 2017.
[3] «ASTM C1673-10ae1 Standard Terminology of C26.10 Nondestructive Assay Methods,» ASTM International, West Consshohocken, PA, 2010.
[4] M. Nizovtsev, S. Stankus, A. Sterlyagov, V. Terekhov y R. Khairulin, «Determination of moisture diffusivity in porous materials using gamma-method,» International Journal of Heat and Mass Transfer, vol. 51, nº 17, pp. 4161-4167, 2008.
[5] A. C. Moreira, C. R. Appoloni, M. V. Oliveira y C. P. Fernandes, «Structural Characterization of Titanium Porous Foams by Gamma Rays Transmission and X Ray Microtomography,» de International Nuclear Atlantic Conference, Santos, SP. Brazil, 2007.
[6] G. Villain y M. Thiery, «Gammadensimetry: A method to determine drying and carbonation profiles in concrete,» NDT&E International, vol. 39, pp. 328-337, 2006.
[7] «ASTM C1718-10 Standard Test Method for Nondestructive Assay of Radioactive Material by Tomographic Gamma Scanning,» ASTM International, West Conshohocken, PA, 2010.
[8] «ASTM C1592/C1592M-09 Standard Guide for Making Quality Nondestructive Assay Measurements,» ASTM International, West Conshohocken, PA, 2009.
[9] National Instruments, LabVIEW™ User Manual, Austin, Texas: © 1992–2003 National Instruments Corporation, April, 2003.
[10] ASM Aerospace Specification Metals Inc., «Aluminum 6061-T6 - ASM Material Data Sheet,» MatWeb, LLC, 2017. [En línea]. Available: http://asm.matweb.com/search/SpecificMaterial.asp?bassnum=ma6061t6. [Último acceso: 08 Agosto 2017].
[11] J. H. Hubbell and S. M. Seltzer, «X-Ray Mass Attenuation Coefficients,» NIST National Institute of Standards and Technology, 30 Junio 2017. [En línea]. Available: https://www.nist.gov/pml/x-ray-mass-attenuation-coefficients. [Último acceso: 08 Agosto 2017]